Triaxial Fluxgate Magnetometer Tri-Mag 225C25X 0505/1515 and Tri-Mag 225C25Z 0505/1515

Features
Data and Performance
Diagrams
Principle
Downloads
Features
  • Non-magnetic hermetic seal packaging for fluxgate circuit
  • Electronic and fluxgate are integrated in one non-magnetic packaging
  • Advanced SOI electronics wafer successfully achieving -60℃ to 225℃
  • Internal temperature sensor for thermal modeling from -50℃ to 200℃
  • Three orthogonal measurement coils together with one permeable core
  • Three components to be measured simultaneously at one point in space
  • Installed along the X-axis or Z-axis direction, suitable for MWD and LWD
Data and Performance

Diagrams
Configuration Drawings

Electrical wiring diagram

Applications
Downhole Directional Sensor Module

 

Principle
The permeable and saturation of magnetic core is excited by the drive circuit. In the permeable state, the magnetic flux is focused to the magnetic core; there is more magnetic flux in the inner of the measurement coils than the outer. In the saturation state, the magnetic flux is diverged from the inner of measurement coils. The magnetic core works as a gate for three measurement coils, which open or close the magnetic flux into inner of each coil, then change the amount of the magnetic flux, and induce a voltage pulse in each coil. In each exciting cycle, the gate state has changed four times, that will induce four voltage pulses: PICKUP1, PICKUP2, PICKUP3 and PICKUP4.

These voltage pulses from each measurement coil have been sampled and integrated in the circuit, accordance to the integration voltage the circuit supply feedback currents respectively to each of the measurement coils, which produce magnetic field completely offset the magnetic field ,as the result, the magnetic field in the inner of three measurement coils is zero. The triaxial fluxgate is closed-loop measuring, each of three feedback currents is exactly equal to each of three axis magnetic field components.
The TRI-MAG225C25X allows the three magnetic field components simultaneously to be measured at the same point in space. Non-magnetic hermetic seal packaging is studied, and three sensitive axies and related circuits are integrated into one housing, and directly output three voltages. The TRI-MAG225C25X adapted SOI (Silicon On Insulator) electronics wafer, hybrid circuit technology and hermetic seal packaging. The TRI-MAG225C25X is able to work in wide temperature range (-50℃ to +200℃) and to survive over 2000 hours.

Calibration coefficient after thermal modeling
The temperature range of thermal modeling is -50℃ to +200℃, the axis alignment is thermal modeled by zero order polynomial, the bias and scale factor are thermal modeled by third order polynomial. each unit will be calibrated before shipping and build up its own calibration coefficients file. To get the accurate X, Y, Z axis component the triaxial fluxgate output voltage shall be calculated with the calibration coefficient.
The accurate magnetic field is obtained with output voltage and calibration coefficient according to following formulas: